Defibrillator |
Operation
method |
Semi-automatic heart shock machine |
Output
waveform |
e-cube biphasic |
Output energy |
200J at 50Ω for adults/ 50J at 50Ω for children |
Appearance |
Size |
Width 260 × Length 256 × Height 69.5mm |
Weight |
Approx. 2.4kg
(Battery pack, pad included) |
Control
system,
Status
display,
CPR
detection |
Control System |
Power button, i-button,
Shock button, adult/child selector switch |
Status display
LCD |
Equipment status, battery level, and pad status display |
CPR
Detection
indicator |
Indicates whether CPR is performed, and whether CPR is correct by light and blink. |
Battery |
type |
12V DC, 4.2Ah LiMnO2 |
Capacity |
At least 200 electric shocks
at first use
approved or continuous
operation 8 hours
available
(At room temperature) |
usage
environment |
Temperature range 0℃ ~ 43℃
(32℉ ~109℉) |
Storage
environment |
Temperature range 0℃ ~ 43℃
(32℉ ~109℉) |
electrocardiogram
analysis
system |
Function |
Analyzes the patient’s impedance and electrocardiogram, and analyzes the rhythm of
the need for defibrillation |
Impedance |
25Ω ~175Ω |
Data storage
and
transmission |
Data storage |
5 data storage in internal memory (max. 3 hours per datum) |
Infrared
communication |
Communication with PC using infrared communication |
SD Card |
ECG data and event data in internal memory can be copied to check them through PC software (CU-Expert). |
Automatic/
Manual
Self-diagnosis
possible |
Automatic
Self-diagnosis |
Power and real-time self-diagnosis test,
Daily, weekly, monthly self-diagnosis test |
Manual
Selfdiagnosis |
Self-diagnosis test when
the user inserts the battery Pack |
Pad |
For adults |
Electrode area: 110㎠
Cable Length: Total 120cm |
For
children |
Electrode area: 50㎠
Cable Length: Total 120cm |